Refine
Has Fulltext
- no (3)
Document Type
- Article (3)
Language
- English (3)
Is part of the Bibliography
- yes (3) (remove)
Keywords
- X-ray diffraction (1)
- acoustic waves (1)
- epitaxial layers (1)
- phonon dispersion relations (1)
- terahertz waves (1)
- thin film devices (1)
Institute
We excite an epitaxial SrRuO3 thin film transducer by a pulse train of ultrashort laser pulses, launching coherent sound waves into the underlying SrTiO3 substrate. Synchrotron-based x-ray diffraction (XRD) data exhibiting separated sidebands to the substrate peak evidence the excitation of a quasi-monochromatic phonon wavepacket with sub-THz central frequency. The frequency and bandwidth of this sound pulse can be controlled by the optical pulse train. We compare the experimental data to combined lattice dynamics and dynamical XRD simulations to verify the coherent phonon dynamics. In addition, we observe a lifetime of 130 ps of such sub-THz phonons in accordance with the theory.
We implemented an experimental scheme for ultrafast x-ray diffraction at storage rings based on a laser-driven Bragg-switch that shortens the x-ray pulses emitted from an undulator. The increased time-resolution is demonstrated by observing changes of intensity, position and width of the diffraction peaks of a La(0.7)Sr(0.3)MnO(3)/SrTiO(3) superlattice sample after optical excitation, i.e., by quantitatively measuring the propagation of an expansion wave through the sample. These experimental transients with timescales of 35 to 60 ps evidence a reduction of the x-ray pulse duration by a factor of two.
We employ the ultrafast response of a 15.4 nm thin SrRuO3 layer grown epitaxially on a SrTiO3 substrate to perform time-domain sampling of an x-ray pulse emitted from a synchrotron storage ring. Excitation of the sample with an ultrashort laser pulse triggers coherent expansion and compression waves in the thin layer, which turn the diffraction efficiency on and off at a fixed Bragg angle during 5 ps. This is significantly shorter than the duration of the synchrotron x-ray pulse of 100 ps. Cross-correlation measurements of the ultrafast sample response and the synchrotron x-ray pulse allow to reconstruct the x-ray pulse shape.