TY - GEN A1 - Brinkmann, Pia A1 - Köllner, Nicole A1 - Merk, Sven A1 - Beitz, Toralf A1 - Altenberger, Uwe A1 - Löhmannsröben, Hans-Gerd T1 - Comparison of handheld and echelle spectrometer to assess copper in ores by means of laser-induced breakdown spectroscopy (LIBS) T2 - Zweitveröffentlichungen der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe N2 - Its properties make copper one of the world’s most important functional metals. Numerous megatrends are increasing the demand for copper. This requires the prospection and exploration of new deposits, as well as the monitoring of copper quality in the various production steps. A promising technique to perform these tasks is Laser Induced Breakdown Spectroscopy (LIBS). Its unique feature, among others, is the ability to measure on site without sample collection and preparation. In this work, copper-bearing minerals from two different deposits are studied. The first set of field samples come from a volcanogenic massive sulfide (VMS) deposit, the second part from a stratiform sedimentary copper (SSC) deposit. Different approaches are used to analyze the data. First, univariate regression (UVR) is used. However, due to the strong influence of matrix effects, this is not suitable for the quantitative analysis of copper grades. Second, the multivariate method of partial least squares regression (PLSR) is used, which is more suitable for quantification. In addition, the effects of the surrounding matrices on the LIBS data are characterized by principal component analysis (PCA), alternative regression methods to PLSR are tested and the PLSR calibration is validated using field samples. T3 - Zweitveröffentlichungen der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe - 1311 KW - LIBS KW - copper-bearing minerals KW - UVR KW - PCA KW - PLSR Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:kobv:517-opus4-584742 SN - 1866-8372 IS - 1311 ER - TY - JOUR A1 - Brinkmann, Pia A1 - Köllner, Nicole A1 - Merk, Sven A1 - Beitz, Toralf A1 - Altenberger, Uwe A1 - Löhmannsröben, Hans-Gerd T1 - Comparison of handheld and echelle spectrometer to assess copper in ores by means of laser-induced breakdown spectroscopy (LIBS) JF - Minerals N2 - Its properties make copper one of the world’s most important functional metals. Numerous megatrends are increasing the demand for copper. This requires the prospection and exploration of new deposits, as well as the monitoring of copper quality in the various production steps. A promising technique to perform these tasks is Laser Induced Breakdown Spectroscopy (LIBS). Its unique feature, among others, is the ability to measure on site without sample collection and preparation. In this work, copper-bearing minerals from two different deposits are studied. The first set of field samples come from a volcanogenic massive sulfide (VMS) deposit, the second part from a stratiform sedimentary copper (SSC) deposit. Different approaches are used to analyze the data. First, univariate regression (UVR) is used. However, due to the strong influence of matrix effects, this is not suitable for the quantitative analysis of copper grades. Second, the multivariate method of partial least squares regression (PLSR) is used, which is more suitable for quantification. In addition, the effects of the surrounding matrices on the LIBS data are characterized by principal component analysis (PCA), alternative regression methods to PLSR are tested and the PLSR calibration is validated using field samples. KW - LIBS KW - copper-bearing minerals KW - UVR KW - PCA KW - PLSR Y1 - 2023 U6 - https://doi.org/10.3390/min13010113 SN - 2075-163X VL - 13 IS - 1 PB - MDPI CY - Basel ER - TY - JOUR A1 - Erler, Alexander A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd A1 - Grothusheitkamp, Daniela A1 - Kunz, Thomas A1 - Methner, Frank-Jürgen T1 - Characterization of volatile metabolites formed by molds on barley by mass and ion mobility spectrometry JF - Journal of mass spectrometr N2 - The contamination of barley by molds on the field or in storage leads to the spoilage of grain and the production of mycotoxins, which causes major economic losses in malting facilities and breweries. Therefore, on-site detection of hidden fungus contaminations in grain storages based on the detection of volatile marker compounds is of high interest. In this work, the volatile metabolites of 10 different fungus species are identified by gas chromatography (GC) combined with two complementary mass spectrometric methods, namely, electron impact (EI) and chemical ionization at atmospheric pressure (APCI)-mass spectrometry (MS). The APCI source utilizes soft X-radiation, which enables the selective protonation of the volatile metabolites largely without side reactions. Nearly 80 volatile or semivolatile compounds from different substance classes, namely, alcohols, aldehydes, ketones, carboxylic acids, esters, substituted aromatic compounds, alkenes, terpenes, oxidized terpenes, sesquiterpenes, and oxidized sesquiterpenes, could be identified. The profiles of volatile and semivolatile metabolites of the different fungus species are characteristic of them and allow their safe differentiation. The application of the same GC parameters and APCI source allows a simple method transfer from MS to ion mobility spectrometry (IMS), which permits on-site analyses of grain stores. Characterization of IMS yields limits of detection very similar to those of APCI-MS. Accordingly, more than 90% of the volatile metabolites found by APCI-MS were also detected in IMS. In addition to different fungus genera, different species of one fungus genus could also be differentiated by GC-IMS. KW - APCI KW - fungus KW - gas chromatography KW - ion mobility spectrometry KW - mass KW - spectrometry KW - mold KW - soft X-ray Y1 - 2020 U6 - https://doi.org/10.1002/jms.4501 SN - 1076-5174 SN - 1096-9888 VL - 55 IS - 5 SP - 1 EP - 10 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Prüfert, Chris A1 - Urban, Raphael David A1 - Fischer, Tillmann Georg A1 - Villatoro, José Andrés A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Belder, Detlev A1 - Zeitler, Kirsten A1 - Löhmannsröben, Hans-Gerd T1 - In situ monitoring of photocatalyzed isomerization reactions on a microchip flow reactor by IR-MALDI ion mobility spectrometry JF - Analytical and bioanalytical chemistry : a merger of Fresenius' journal of analytical chemistry, Analusis and Quimica analitica N2 - The visible-light photocatalyticE/Zisomerization of olefins can be mediated by a wide spectrum of triplet sensitizers (photocatalysts). However, the search for the most efficient photocatalysts through screenings in photo batch reactors is material and time consuming. Capillary and microchip flow reactors can accelerate this screening process. Combined with a fast analytical technique for isomer differentiation, these reactors can enable high-throughput analyses. Ion mobility (IM) spectrometry is a cost-effective technique that allows simple isomer separation and detection on the millisecond timescale. This work introduces a hyphenation method consisting of a microchip reactor and an infrared matrix-assisted laser desorption ionization (IR-MALDI) ion mobility spectrometer that has the potential for high-throughput analysis. The photocatalyzedE/Zisomerization of ethyl-3-(pyridine-3-yl)but-2-enoate (E-1) as a model substrate was chosen to demonstrate the capability of this device. Classic organic triplet sensitizers as well as Ru-, Ir-, and Cu-based complexes were tested as catalysts. The ionization efficiency of theZ-isomer is much higher at atmospheric pressure which is due to a higher proton affinity. In order to suppress proton transfer reactions by limiting the number of collisions, an IM spectrometer working at reduced pressure (max. 100 mbar) was employed. This design reduced charge transfer reactions and allowed the quantitative determination of the reaction yield in real time. Among 14 catalysts tested, four catalysts could be determined as efficient sensitizers for theE/Zisomerization of ethyl cinnamate derivativeE-1. Conversion rates of up to 80% were achieved in irradiation time sequences of 10 up to 180 s. With respect to current studies found in the literature, this reduces the acquisition times from several hours to only a few minutes per scan. KW - microchip KW - reaction monitoring KW - IR-MALDI KW - ion mobility spectrometry KW - photochemistry KW - photocatalysis KW - Olefin isomerization Y1 - 2020 U6 - https://doi.org/10.1007/s00216-020-02923-y SN - 1618-2642 SN - 1618-2650 VL - 412 IS - 28 SP - 7899 EP - 7911 PB - Springer CY - Heidelberg ER - TY - JOUR A1 - Prüfert, Christian A1 - Villatoro Leal, José Andrés A1 - Zühlke, Martin A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd T1 - Liquid phase IR-MALDI and differential mobility analysis of nano- and sub-micron particles JF - Physical chemistry, chemical physics : a journal of European Chemical Societies N2 - Infrared matrix-assisted desorption and ionization (IR-MALDI) enables the transfer of sub-micron particles (sMP) directly from suspensions into the gas phase and their characterization with differential mobility (DM) analysis. A nanosecond laser pulse at 2940 nm induces a phase explosion of the aqueous phase, dispersing the sample into nano- and microdroplets. The particles are ejected from the aqueous phase and become charged. Using IR-MALDI on sMP of up to 500 nm in diameter made it possible to surpass the 100 nm size barrier often encountered when using nano-electrospray for ionizing supramolecular structures. Thus, the charge distribution produced by IR-MALDI could be characterized systematically in the 50-500 nm size range. Well-resolved signals for up to octuply charged particles were obtained in both polarities for different particle sizes, materials, and surface modifications spanning over four orders of magnitude in concentrations. The physicochemical characterization of the IR-MALDI process was done via a detailed analysis of the charge distribution of the emerging particles, qualitatively as well as quantitatively. The Wiedensohler charge distribution, which describes the evolution of particle charging events in the gas phase, and a Poisson-derived charge distribution, which describes the evolution of charging events in the liquid phase, were compared with one another with respect to how well they describe the experimental data. Although deviations were found in both models, the IR-MALDI charging process seems to resemble a Poisson-like charge distribution mechanism, rather than a bipolar gas phase charging one. Y1 - 2022 U6 - https://doi.org/10.1039/d1cp04196g SN - 1463-9076 SN - 1463-9084 VL - 24 IS - 4 SP - 2275 EP - 2286 PB - Royal Society of Chemistry CY - Cambridge ER - TY - JOUR A1 - Morgner, Frank A1 - Bennemann, Mark A1 - Cywiński, Piotr J. A1 - Kollosche, Matthias A1 - Górski, Krzysztof A1 - Pietraszkiewicz, Marek A1 - Geßner, André A1 - Löhmannsröben, Hans-Gerd T1 - Elastic FRET sensors for contactless pressure measurement JF - RSC Advances : an international journal to further the chemical sciences N2 - Contactless pressure monitoring based on Forster resonance energy transfer between donor/acceptor pairs immobilized within elastomers is demonstrated. The donor/acceptor energy transfer is employed by dispersing terbium(III) tris[(2-hydroxybenzoyl)-2-aminoethyl] amine complex (LLC, donor) and CdSe/ZnS quantum dots (QD655, acceptor) in styrene-ethylene/buthylene-styrene (SEBS) and poly(dimethylsiloxane) (PDMS). The continuous monitoring of QD luminescence showed a reversible intensity change as the pressure signal is alternated between two stable states indicating a pressure sensitivity of 6350 cps kPa(-1). Time-resolved measurements show the pressure impact on the FRET signal due to an increase of decay time (270 ms up to 420 ms) for the donor signal and parallel drop of decay time (170 mu s to 155 mu s) for the acceptor signal as the net pressure applied. The LLC/QD655 sensors enable a contactless readout as well as space resolved monitoring to enable miniaturization towards smaller integrated stretchable opto-electronics. Elastic FRET sensors can potentially lead to developing profitable analysis systems capable to outdo conventional wired electronic systems (inductive, capacitive, ultrasonic and photoelectric sensors) especially for point-of-care diagnostics, biological monitoring required for wearable electronics. Y1 - 2017 U6 - https://doi.org/10.1039/c7ra06379b SN - 2046-2069 VL - 7 SP - 50578 EP - 50583 PB - RSC Publishing CY - Cambridge ER - TY - GEN A1 - Rethfeldt, Nina A1 - Brinkmann, Pia A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Köllner, Nicole A1 - Altenberger, Uwe A1 - Löhmannsröben, Hans-Gerd T1 - Detection of Rare Earth Elements in Minerals and Soils by Laser-Induced Breakdown Spectroscopy (LIBS) Using Interval PLS T2 - Zweitveröffentlichungen der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe N2 - The numerous applications of rare earth elements (REE) has lead to a growing global demand and to the search for new REE deposits. One promising technique for exploration of these deposits is laser-induced breakdown spectroscopy (LIBS). Among a number of advantages of the technique is the possibility to perform on-site measurements without sample preparation. Since the exploration of a deposit is based on the analysis of various geological compartments of the surrounding area, REE-bearing rock and soil samples were analyzed in this work. The field samples are from three European REE deposits in Sweden and Norway. The focus is on the REE cerium, lanthanum, neodymium and yttrium. Two different approaches of data analysis were used for the evaluation. The first approach is univariate regression (UVR). While this approach was successful for the analysis of synthetic REE samples, the quantitative analysis of field samples from different sites was influenced by matrix effects. Principal component analysis (PCA) can be used to determine the origin of the samples from the three deposits. The second approach is based on multivariate regression methods, in particular interval PLS (iPLS) regression. In comparison to UVR, this method is better suited for the determination of REE contents in heterogeneous field samples. View Full-Text T3 - Zweitveröffentlichungen der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe - 1254 KW - LIBS KW - rare earth elements KW - minerals KW - PCA KW - iPLS regression Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:kobv:517-opus4-557469 SN - 1866-8372 SP - 1 EP - 17 PB - Universitätsverlag Potsdam CY - Potsdam ER - TY - JOUR A1 - Rethfeldt, Nina A1 - Brinkmann, Pia A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Köllner, Nicole A1 - Altenberger, Uwe A1 - Löhmannsröben, Hans-Gerd T1 - Detection of Rare Earth Elements in Minerals and Soils by Laser-Induced Breakdown Spectroscopy (LIBS) Using Interval PLS JF - Minerals N2 - The numerous applications of rare earth elements (REE) has lead to a growing global demand and to the search for new REE deposits. One promising technique for exploration of these deposits is laser-induced breakdown spectroscopy (LIBS). Among a number of advantages of the technique is the possibility to perform on-site measurements without sample preparation. Since the exploration of a deposit is based on the analysis of various geological compartments of the surrounding area, REE-bearing rock and soil samples were analyzed in this work. The field samples are from three European REE deposits in Sweden and Norway. The focus is on the REE cerium, lanthanum, neodymium and yttrium. Two different approaches of data analysis were used for the evaluation. The first approach is univariate regression (UVR). While this approach was successful for the analysis of synthetic REE samples, the quantitative analysis of field samples from different sites was influenced by matrix effects. Principal component analysis (PCA) can be used to determine the origin of the samples from the three deposits. The second approach is based on multivariate regression methods, in particular interval PLS (iPLS) regression. In comparison to UVR, this method is better suited for the determination of REE contents in heterogeneous field samples. View Full-Text KW - LIBS KW - rare earth elements KW - minerals KW - PCA KW - iPLS regression Y1 - 2021 U6 - https://doi.org/10.3390/min11121379 SN - 2075-163X VL - 11 SP - 1 EP - 17 PB - MDPI CY - Basel, Schweiz ER - TY - JOUR A1 - Michalik-Onichimowska, Aleksandra A1 - Beitz, Toralf A1 - Panne, Ulrich A1 - Löhmannsröben, Hans-Gerd A1 - Riedel, Jens T1 - Microsecond mid-infrared laser pulses for atmospheric pressure laser ablation/ionization of liquid samples JF - Sensors and actuators : B, Chemical N2 - In many laser based ionization techniques with a subsequent drift time separation, the laser pulse generating the ions is considered as the start time to. Therefore, an accurate temporal definition of this event is crucial for the resolution of the experiments. In this contribution, the laser induced plume dynamics of liquids evaporating into atmospheric pressure are visualized for two distinctively different laser pulse widths, Delta t = 6 nanoseconds and Delta tau = 280 microseconds. For ns-pulses the expansion of the generated vapour against atmospheric pressure is found to lead to turbulences inside the gas phase. This results in spatial and temporal broadening of the nascent clouds. A more equilibrated expansion, without artificial smearing of the temporal resolution can, in contrast, be observed to follow mu s-pulse excitation. This leads to the counterintuitive finding that longer laser pulses results in an increased temporal vapour formation definition. To examine if this fume expansion also eventually results in a better definition of ion formation, the nascent vapour plumes were expanded into a linear drift tube ion mobility spectrometer (IMS). This time resolved detection of ion formation corroborates the temporal broadening caused by collisional impeding of the supersonic expansion at atmospheric pressure and the overall better defined ion formation by evaporation with long laser pulses. A direct comparison of the observed results strongly suggests the coexistence of two individual ion formation mechanisms that can be specifically addressed by the use of appropriate laser sources. KW - Laser ablation KW - Ion mobility spectrometry KW - Pulse duration KW - Plume KW - Ionization Y1 - 2016 U6 - https://doi.org/10.1016/j.snb.2016.06.155 SN - 0925-4005 VL - 238 SP - 298 EP - 305 PB - Elsevier CY - Lausanne ER - TY - GEN A1 - Ellis, S. C. A1 - Bauer, S. A1 - Bacigalupo, C. A1 - Bland-Hawthorn, J. A1 - Bryant, J. J. A1 - Case, S. A1 - Content, R. A1 - Fechner, T. A1 - Giannone, D. A1 - Haynes, R. A1 - Hernandez, E. A1 - Horton, A. J. A1 - Klauser, U. A1 - Lawrence, J. S. A1 - Leon-Saval, S. G. A1 - Lindley, E. A1 - Löhmannsröben, Hans-Gerd A1 - Min, S. -S. A1 - Pai, N. A1 - Roth, M. A1 - Shortridge, K. A1 - Waller, L. A1 - Xavier, Pascal A1 - Zhelem, Ross T1 - PRAXIS: an OH suppression optimised near infrared spectrograph T2 - Ground-based and Airborne Instrumentation for Astronomy VII N2 - The problem of atmospheric emission from OH molecules is a long standing problem for near-infrared astronomy. PRAXIS is a unique spectrograph which is fed by fibres that remove the OH background and is optimised specifically to benefit from OH-Suppression. The OH suppression is achieved with fibre Bragg gratings, which were tested successfully on the GNOSIS instrument. PRAXIS uses the same fibre Bragg gratings as GNOSIS in its first implementation, and will exploit new, cheaper and more efficient, multicore fibre Bragg gratings in the second implementation. The OH lines are suppressed by a factor of similar to 1000, and the expected increase in the signal-to-noise in the interline regions compared to GNOSIS is a factor of similar to 9 with the GNOSIS gratings and a factor of similar to 17 with the new gratings. PRAXIS will enable the full exploitation of OH suppression for the first time, which was not achieved by GNOSIS (a retrofit to an existing instrument that was not OH-Suppression optimised) due to high thermal emission, low spectrograph transmission and detector noise. PRAXIS has extremely low thermal emission, through the cooling of all significantly emitting parts, including the fore-optics, the fibre Bragg gratings, a long length of fibre, and the fibre slit, and an optical design that minimises leaks of thermal emission from outside the spectrograph. PRAXIS has low detector noise through the use of a Hawaii-2RG detector, and a high throughput through a efficient VPH based spectrograph. PRAXIS will determine the absolute level of the interline continuum and enable observations of individual objects via an IFU. In this paper we give a status update and report on acceptance tests. KW - Near infrared KW - spectroscopy KW - OH suppression KW - astrophotonics KW - fibre Bragg gratings Y1 - 2018 SN - 978-1-5106-1958-6 U6 - https://doi.org/10.1117/12.2311898 SN - 0277-786X SN - 1996-756X VL - 10702 PB - SPIE-INT Soc Optical Engineering CY - Bellingham ER - TY - JOUR A1 - Zühlke, Martin A1 - Meiling, Till Thomas A1 - Roder, Phillip A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Bald, Ilko A1 - Löhmannsröben, Hans-Gerd A1 - Janßen, Traute A1 - Erhard, Marcel A1 - Repp, Alexander T1 - Photodynamic inactivation of E. coli bacteria via carbon nanodots JF - ACS omega / American Chemical Society N2 - The increasing development of antibiotic resistance in bacteria has been a major problem for years, both in human and veterinary medicine. Prophylactic measures, such as the use of vaccines, are of great importance in reducing the use of antibiotics in livestock. These vaccines are mainly produced based on formaldehyde inactivation. However, the latter damages the recognition elements of the bacterial proteins and thus could reduce the immune response in the animal. An alternative inactivation method developed in this work is based on gentle photodynamic inactivation using carbon nanodots (CNDs) at excitation wavelengths λex > 290 nm. The photodynamic inactivation was characterized on the nonvirulent laboratory strain Escherichia coli K12 using synthesized CNDs. For a gentle inactivation, the CNDs must be absorbed into the cytoplasm of the E. coli cell. Thus, the inactivation through photoinduced formation of reactive oxygen species only takes place inside the bacterium, which means that the outer membrane is neither damaged nor altered. The loading of the CNDs into E. coli was examined using fluorescence microscopy. Complete loading of the bacterial cells could be achieved in less than 10 min. These studies revealed a reversible uptake process allowing the recovery and reuse of the CNDs after irradiation and before the administration of the vaccine. The success of photodynamic inactivation was verified by viability assays on agar. In a homemade flow photoreactor, the fastest successful irradiation of the bacteria could be carried out in 34 s. Therefore, the photodynamic inactivation based on CNDs is very effective. The membrane integrity of the bacteria after irradiation was verified by slide agglutination and atomic force microscopy. The method developed for the laboratory strain E. coli K12 could then be successfully applied to the important avian pathogens Bordetella avium and Ornithobacterium rhinotracheale to aid the development of novel vaccines. KW - Bacteria KW - Genetics KW - Fluorescence KW - Photodynamics KW - Irradiation Y1 - 2021 U6 - https://doi.org/10.1021/acsomega.1c01700 SN - 2470-1343 VL - 6 IS - 37 SP - 23742 EP - 23749 PB - ACS Publications CY - Washington, DC ER - TY - JOUR A1 - Meiling, Till Thomas A1 - Cywinski, Piotr J. A1 - Löhmannsröben, Hans-Gerd T1 - Two-Photon excitation fluorescence spectroscopy of quantum dots BT - photophysical properties and application in bioassays JF - The journal of physical chemistry : C, Nanomaterials and interfaces N2 - The applications of quantum dots (QDs) in two-photon (2P) excitation applications demand reliable data about their 2P absorption (2PA) cross sections (sigma(2PA)). In the present study, sigma(2PA) values have been determined for a series of commercial colloidal CdSe/ZnS QDs and CdSeTe/ZnS QDs in aqueous media. For the first time for these QDs, the sigma(2PA) values have been determined over a wide spectral range, that is, between 720 and 900 nm, and are compared to the extinction coefficient (epsilon) values obtained under one-photon (1P) excitation. Furthermore, we present a QD in combination with an organic dye in a biotin-streptavidin Forster resonance energy transfer bioassay under 1P and 2P excitation. The results for the bioassay under 2P excitation are compared to those obtained under 1P excitation. The results demonstrate that in the case of the 2P excitation, higher sensitivity can be achieved because of an improved signal-to-noise ratio. Y1 - 2018 U6 - https://doi.org/10.1021/acs.jpcc.7b12345 SN - 1932-7447 SN - 1932-7455 VL - 122 IS - 17 SP - 9641 EP - 9647 PB - American Chemical Society CY - Washington ER - TY - JOUR A1 - Erler, Alexander A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd A1 - Grothusheitkamp, Daniela A1 - Kunz, T. A1 - Methner, Frank-Jürgen T1 - Detection of volatile organic compounds in the headspace above mold fungi by GC-soft X-radiation-based APCI-MS JF - Journal of mass spectrometr N2 - Mold fungi on malting barley grains cause major economic loss in malting and brewery facilities. Possible proxies for their detection are volatile and semivolatile metabolites. Among those substances, characteristic marker compounds have to be identified for a confident detection of mold fungi in varying surroundings. The analytical determination is usually performed through passive sampling with solid phase microextraction, gas chromatographic separation, and detection by electron ionization mass spectrometry (EI-MS), which often does not allow a confident determination due to the absence of molecular ions. An alternative is GC-APCI-MS, generally, allowing the determination of protonated molecular ions. Commercial atmospheric pressure chemical ionization (APCI) sources are based on corona discharges, which are often unspecific due to the occurrence of several side reactions and produce complex product ion spectra. To overcome this issue, an APCI source based on soft X-radiation is used here. This source facilitates a more specific ionization by proton transfer reactions only. In the first part, the APCI source is characterized with representative volatile fungus metabolites. Depending on the proton affinity of the metabolites, the limits of detection are up to 2 orders of magnitude below those of EI-MS. In the second part, the volatile metabolites of the mold fungus species Aspergillus, Alternaria, Fusarium, and Penicillium are investigated. In total, 86 compounds were found with GC-EI/APCI-MS. The metabolites identified belong to the substance classes of alcohols, aldehydes, ketones, carboxylic acids, esters, substituted aromatic compounds, terpenes, and sesquiterpenes. In addition to substances unspecific for the individual fungus species, characteristic patterns of metabolites, allowing their confident discrimination, were found for each of the 4 fungus species. Sixty-seven of the 86 metabolites are detected by X-ray-based APCI-MS alone. The discrimination of the fungus species based on these metabolites alone was possible. Therefore, APCI-MS in combination with collision induced dissociation alone could be used as a supervision method for the detection of mold fungi. KW - APCI KW - gas chromatography KW - mass spectrometry KW - mold fungi KW - soft X-radiation KW - volatile organic compounds Y1 - 2018 U6 - https://doi.org/10.1002/jms.4210 SN - 1076-5174 SN - 1096-9888 VL - 53 IS - 10 SP - 911 EP - 920 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Zühlke, Martin A1 - Sass, Stephan A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd T1 - Real-Time Reaction Monitoring of an Organic Multistep Reaction by Electrospray Ionization-Ion Mobility Spectrometry JF - ChemPlusChem N2 - The capability of electrospray ionization (ESI)-ion mobility (IM) spectrometry for reaction monitoring is assessed both as a stand-alone real-time technique and in combination with HPLC. A three-step chemical reaction, consisting of a Williamson ether synthesis followed by a hydrogenation and an N-alkylation step, is chosen for demonstration. Intermediates and products are determined with a drift time to mass-per-charge correlation. Addition of an HPLC column to the setup increases the separation power and allows the determination of further species. Monitoring of the intensities of the various species over the reaction time allows the detection of the end of reaction, determination of the rate-limiting step, observation of the system response in discontinuous processes, and optimization of the mass ratios of the starting materials. However, charge competition in ESI influences the quantitative detection of substances in the reaction mixture. Therefore, two different methods are investigated, which allow the quantification and investigation of reaction kinetics. The first method is based on the pre-separation of the compounds on an HPLC column and their subsequent individual detection in the ESI-IM spectrometer. The second method involves an extended calibration procedure, which considers charge competition effects and facilitates nearly real-time quantification. KW - electrospray ionization KW - HPLC KW - ion mobility spectrometry KW - reaction mechanisms KW - reaction monitoring Y1 - 2017 U6 - https://doi.org/10.1002/cplu.201700296 SN - 2192-6506 VL - 82 SP - 1266 EP - 1273 PB - Wiley-VCH CY - Weinheim ER - TY - GEN A1 - Eich, Susanne A1 - Schmälzlin, Elmar A1 - Löhmannsröben, Hans-Gerd T1 - Distributed fiber optical sensing of oxygen with optical time domain reflectometry T2 - Postprints der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe N2 - In many biological and environmental applications spatially resolved sensing of molecular oxygen is desirable. A powerful tool for distributed measurements is optical time domain reflectometry (OTDR) which is often used in the field of telecommunications. We combine this technique with a novel optical oxygen sensor dye, triangular-[4] phenylene (TP), immobilized in a polymer matrix. The TP luminescence decay time is 86 ns. The short decay time of the sensor dye is suitable to achieve a spatial resolution of some meters. In this paper we present the development and characterization of a reflectometer in the UV range of the electromagnetic spectrum as well as optical oxygen sensing with different fiber arrangements. T3 - Zweitveröffentlichungen der Universität Potsdam : Mathematisch-Naturwissenschaftliche Reihe - 1085 KW - OTDR KW - optical sensing KW - molecular oxygen KW - triangular-[4] phenylene Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:kobv:517-opus4-476659 SN - 1866-8372 IS - 1085 ER - TY - JOUR A1 - Zühlke, Martin A1 - Zenichowski, Karl A1 - Riebe, Daniel A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd T1 - Subambient pressure electrospray ionization ion mobility spectrometry JF - International journal for ion mobility spectrometry : official publication of the International Society for Ion Mobility Spectrometry N2 - The pressure dependence of sheath gas assisted electrospray ionization (ESI) was investigated based on two complementary experimental setups, namely an ESI-ion mobility (IM) spectrometer and an ESI capillary - Faraday plate setup housed in an optically accessible vacuum chamber. The ESI-IM spectrometer is capable of working in the pressure range between 300 and 1000 mbar. Another aim was the assessment of the analytical capabilities of a subambient pressure ESI-IM spectrometer. The pressure dependence of ESI was characterized by imaging the electrospray and recording current-voltage (I-U) curves. Qualitatively different behavior was observed in both setups. While the current rises continuously with the voltage in the capillary-plate setup, a sharp increase of the current was measured in the IM spectrometer above a pressure-dependent threshold voltage. The different character can be attributed to the detection of different species in both experiments. In the capillary-plate experiment, a multitude of charged species are detected while only desolvated ions attribute to the IM spectrometer signal. This finding demonstrates the utility of IM spectrometry for the characterization of ESI, since in contrast to the capillary-plate setup, the release of ions from the electrospray droplets can be observed. The I-U curves change significantly with pressure. An important result is the reduction of the maximum current with decreasing pressure. The connected loss of ionization efficiency can be compensated by a more efficient transfer of ions in the IM spectrometer at increased E/N. Thus, similar limits of detection could be obtained at 500 mbar and 1 bar. KW - Ion mobility spectrometry KW - Electrospray ionization KW - Subambient pressure KW - Imaging Y1 - 2017 U6 - https://doi.org/10.1007/s12127-017-0215-x SN - 1435-6163 SN - 1865-4584 VL - 20 SP - 47 EP - 56 PB - Springer CY - Heidelberg ER - TY - JOUR A1 - Khadem, S. M. J. A1 - Hille, Carsten A1 - Löhmannsröben, Hans-Gerd A1 - Sokolov, Igor M. T1 - Spot variation fluorescence correlation spectroscopy by data post-processing JF - Scientific reports N2 - Spot variation fluorescence correlation spectroscopy (SV-FCS) is a variant of the FCS techniques which may give useful information about the structural organisation of the medium in which the diffusion takes place. We show that the same results can be obtained by post-processing the photon count data from ordinary FCS measurements. By using this method, one obtains the fluorescence autocorrelation functions for sizes of confocal volume, which are effectively smaller than that of the initial FCS measurement. The photon counts of the initial experiment are first transformed into smooth intensity trace using kernel smoothing method or to a piecewise-continuous intensity trace using binning and then a non-linear transformation is applied to this trace. The result of this transformation mimics the photon count rate in an experiment performed with a smaller confocal volume. The applicability of the method is established in extensive numerical simulations and directly supported in in-vitro experiments. The procedure is then applied to the diffusion of AlexaFluor647-labeled streptavidin in living cells. Y1 - 2017 U6 - https://doi.org/10.1038/s41598-017-05672-8 SN - 2045-2322 VL - 7 SP - 1 EP - 9 PB - Nature Publ. Group CY - London ER - TY - JOUR A1 - Cywinski, Piotr J. A1 - Pietraszkiewicz, Marek A1 - Maciejczyk, Michal A1 - Gorski, Krzysztof A1 - Hammann, Tommy A1 - Liermann, Konstanze A1 - Paulke, Bernd-Reiner A1 - Löhmannsröben, Hans-Gerd T1 - Total protein concentration quantification using nanobeads with a new highly luminescent terbium(III) complex JF - RSC Advances N2 - Total protein concentration (TPC) is a key parameter in many biochemical experiments and its quantification is often necessary for isolation, separation, and analysis of proteins. A sensitive and rapid nanobead-based TPC quantification assay based on Forster Resonance Energy Transfer (FRET) has been developed. A new, highly luminescent Tb(III) complex has been synthesised and applied as donor in this FRET assay with an organic dye (Cy5) as acceptor. FRET-induced changes in luminescence have been investigated both at donor and acceptor emission wavelength using time-resolved luminescence spectroscopy with time-gated detection. In the assay, the Tb(III) complex and fine-tuned polyglycidyl methacrylate (PGMA) nanobeads ensure that an improvement in sensitivity and background reduction is achieved. Using 40 nm large PGMA nanobeads loaded with the Tb(III) complex, it is possible to determine TPC down to 50 ng mL(-1) in just 10 minutes. Through specific assay components the sensitivity has been improved when compared to existing nanobead-based assays and to currently known commercial methods. Additionally, the assay is relatively insensitive to the presence of contaminants, such as non-ionic detergents commonly found in biological samples. Due to no need for any centrifugal steps, this mix-and-measure bioassay can easily be implemented into routine TPC quantification protocols in biochemical laboratories. Y1 - 2016 U6 - https://doi.org/10.1039/c6ra23207h SN - 2046-2069 VL - 6 SP - 115068 EP - 115073 PB - Royal Society of Chemistry CY - Cambridge ER - TY - JOUR A1 - Riebe, Daniel A1 - Eder, Alexander A1 - Ritschel, Thomas A1 - Beitz, Toralf A1 - Löhmannsröben, Hans-Gerd A1 - Beil, Andreas A1 - Blaschke, Michael A1 - Ludwig, Thomas T1 - Atmospheric pressure chemical ionization of explosives induced by soft X-radiation in ion mobility spectrometry: mass spectrometric investigation of the ionization reactions of drift gasses, dopants and alkyl nitrates JF - Journal of mass spectrometr N2 - A promising replacement for the radioactive sources commonly encountered in ion mobility spectrometers is a miniaturized, energy-efficient photoionization source that produce the reactant ions via soft X-radiation (2.8 keV). In order to successfully apply the photoionization source, it is imperative to know the spectrum of reactant ions and the subsequent ionization reactions leading to the detection of analytes. To that end, an ionization chamber based on the photoionization source that reproduces the ionization processes in the ion mobility spectrometer and facilitates efficient transfer of the product ions into a mass spectrometer was developed. Photoionization of pure gasses and gas mixtures containing air, N-2, CO2 and N2O and the dopant CH2Cl2 is discussed. The main product ions of photoionization are identified and compared with the spectrum of reactant ions formed by radioactive and corona discharge sources on the basis of literature data. The results suggest that photoionization by soft X-radiation in the negative mode is more selective than the other sources. In air, adduct ions of O-2 - with H2O and CO2 were exclusively detected. Traces of CO2 impact the formation of adduct ions of O-2 - and Cl -(upon addition of dopant) and are capable of suppressing them almost completely at high CO2 concentrations. Additionally, the ionization products of four alkyl nitrates (ethylene glycol dinitrate, nitroglycerin, erythritol tetranitrate and pentaerythritol tetranitrate) formed by atmospheric pressure chemical ionization induced by X-ray photoionization in different gasses (air, N-2 and N2O) and dopants (CH2Cl2, C2H5Br and CH3I) are investigated. The experimental studies are complemented by density functional theory calculations of the most important adduct ions of the alkyl nitrates (M) used for their spectrometric identification. In addition to the adduct ions [M + NO3](-) and [M + Cl](-), adduct ions such as [M + N2O2](-), [M + Br](-) and [M+ I](-) were detected, and their gas-phase structures and energetics are investigated by density functional theory calculations. Copyright (C) 2016 John Wiley & Sons, Ltd. KW - ion mobility spectrometry KW - mass spectrometry KW - explosives KW - X-ray KW - photoionization KW - alkyl nitrates Y1 - 2016 U6 - https://doi.org/10.1002/jms.3784 SN - 1076-5174 SN - 1096-9888 VL - 51 SP - 566 EP - 577 PB - Wiley-Blackwell CY - Hoboken ER - TY - JOUR A1 - Khadem, S. M. J. A1 - Hille, Carsten A1 - Löhmannsröben, Hans-Gerd A1 - Sokolov, Igor M. T1 - What information is contained in the fluorescence correlation spectroscopy curves, and where JF - Physical review : E, Statistical, nonlinear and soft matter physics Y1 - 2016 U6 - https://doi.org/10.1103/PhysRevE.94.022407 SN - 2470-0045 SN - 2470-0053 VL - 94 PB - American Physical Society CY - College Park ER -