TY - JOUR A1 - Wilke, Max A1 - Appel, Karen A1 - Vincze, Laszlo A1 - Schmidt, Christian A1 - Borchert, Manuela A1 - Pascarelli, Sakura T1 - A confocal set-up for micro-XRF and XAFS experiments using diamond-anvil cells N2 - A confocal set-up is presented that improves micro-XRF and XAFS experiment with high-pressure e diamond-anvil cells (DACs) In this experiment a probing volume is defined by the focus of the incoming synchrotron radiation beam and that of a polycapillary X-ray half-lens with a very long working distance, which is placed in front of the fluorescence detector This set-up enhances the quality of the fluorescence and XAFS spectra, and thus the sensitivity for detecting elements at low concentrations. It efficiently suppresses signal from outside the sample chamber, which stems from elastic and inelastic scattering of the incoming beam by the diamond anvils as well as from excitation of fluorescence from the body of the DAC Y1 - 2010 UR - https://publishup.uni-potsdam.de/frontdoor/index/index/docId/12504 UR - http://journals.iucr.org/s/journalhomepage.html SN - 0909-0495 ER -